The FP-260 is a highly sensitive toxic gas monitor using a gas detection cassette tape which is almost free of interference from other gases.
Since the detection tape responds only to the chemical reaction of the gas being measured, there is no interference from hydrogen or organic gases.
Its best application then is in monitoirng of the target gas at the outlet of an absorption scrubber, as well as in monitoring of the work environment for personnel safety.
Since the cassette can be replaced so quickly and easily by a one touch cassette method, detection of multiple gases is practical.
Features
Applications
Detectable Gases and Detection Ranges
Gas |
TLV value |
Detection Range |
Resolution |
| Phosphine (PH3) | 300 ppb |
0~900 ppb |
5 |
| Silane (SiH4) | 5 ppb |
0~15 ppm |
0.25 |
| Diborane (B2H6) | 100 ppb |
0~300 ppb |
5 |
| Arsine (AsH3) | 50 ppb |
0~150 ppb |
1 |
| Disilane (Si2H6) | 5 ppm |
0~10 ppm |
0.1 |
| Hydrogen Sulfide (H2S) | 10 ppm |
0~100 ppb |
1 |
| Phosgene (COCl2) | 100 ppb |
0~300 ppb |
2 |
| Chlorine (Cl2) | 0.5 ppm |
0~100 ppb |
1 |
| Hydrogen selenide (H2Se) | 50 ppb |
0~300 ppb |
2 |
Specifications
Model |
FP-260 |
Type |
Desktop type, transportable |
Detection principle |
Photoelectric photometry method using paper tape cassette |
Detection tape |
Cassette type, in VHS style quick change cassette |
Measuring gas and range |
see table above |
Accuracy |
Within (+/- 20% of indication value +1 digit) at constant conditions |
Tape operation time |
1 month (with no alarm) - Tape remainder display and caution - Alarm display at tape end |
Display |
Digital LCD |
Alarm Preset Value |
Warning: 1/6 of full scale Alarm: 1/3 of full scale |
Alarm Display |
Lamp (warning-yellow, alarm-red) and buzzer |
Alarm Output |
Latching (standard) or non latching (option) - Dry contact 1A (warning) latching AC 125V, 0.5A - Dry contact 1A (alarm) latching AC 125V, 0.5A |
Response Time |
Within 60 seconds when exposed to gas concentration of 1.6 times alarm level. |
Self diagnosis |
Sensor trouble, flow fail, warning of tape end, tape exchange or tape damage, system trouble. Lamp and LCD message "C" contact (No voltage, self-diagnosis) rating : AC 125V, 0.5A |
External output signal |
4-20mA, max load resistance 300Ω (standard or Multiplex transmission (option) |
Operating temperature & humidity |
5~35°C (41~95°F), 30~90% RH (non condensing) |
Sampling flow rate |
Approx. 1.0 l/min |
Gas Inlet |
Outer diameter: 6mm, teflon tube with fitting |
Power supply & power |
100VAC, 115 VAC, 220VAC 50/60 Hz (35VA at Measurement, 60VA at feeding tape) |
Dimensions & weight |
Approx. 275(W) x 220(H) x 370(D)mm Approx. 12 kg |
Standard Accessories |
Gas detection cassette tape, Dust filter, Fuse (2A) |